Appascan Max (A Scan)
Technical Specifications
| A-scan Probe | |
| Probe frequency | 10 MHZ |
| Fixation | Internal LED |
| Measurement range | |
| Axial Length | 14.0 mm to 40.0mm |
| ACD | 2.0 mm to 6.0 mm |
| Lens Thickness | 2.0 mm to 7.5mm |
| Measurement Accuracy | ± 0.1mm |
| Measurement Techniques | |
| Method | Immersion / Contact |
| Mode | Auto / Manual |
| Lens Type | Aphakic, Normal, Cataract, Dense Cataract, Pseudophakic (Silicone, PMMA,Others), Silicone Oil |
| IOL Calculation | |
| Formulae | SRK-T, SRK II, Holladay, Binkhorst, Hoffer-Q, Haigis, AME*, and Post Refractive Formulas |
| Display | High resolution 8 inch colour display With Touch screen |
| Control | |
| Storage | 100 patient records, 10 Preset User, 25 IOL configurations |
| Camp Mode | Fast Measurement with IOL calculation in single screen |
| Printer | Fast Thermal printer with Auto cutter |
| Power | |
| Electrical Requirements | 230V AC / 50HZ |
Other Features
- Auto / Manual Gain
- Automatic calculation of standard deviation and average axial length (series of 10 measurements)
- Simultaneous display of 3 different IOL calculations
- RTC shows the Date and Time throughout the life of the instrument.
- Various user Programmable printouts Format
- Measurement possible for Various Velocity of each segment .
- *AME Formula – The new generation programmed software takes care of any Axial Length.
Accessories
- Soundless Motorized Stand
Appascan Max with Pachymeter
Technical Specifications
| ASCAN PROBE | |
| Probe Frequency | 10MHZ |
| Fixation | Internal LED |
| Measurement range | |
| Axial Length | 14.0 mm to 40.0mm |
| ACD | 2.0 mm to 6.0 mm |
| Lens Thickness | 2.0 mm to 7.5mm |
| Measurement Accuracy | + 0.1 mm |
| Measurement Techniques | |
| Method | Immersion / Contact |
| Mode | Auto / Manual |
| Lens Type | Aphakic, Normal, Cataract, Dense Cataract, Pseudophakic, ( Silicone, PMMA, Others ) Silicone Oil |
| IOL Calculation | |
| Formulae | SRK – T, SRK – II, Holladay, Binkhorst, Hoffer-Q, Haigis, AME* and Post Refraction Formulae |
| Display Control | High resolution 8 inch colour display With Touch screen |
| Storage | 100 Patient records, 10 preset users 25 IOL Configuration |
| Camp Mode | Fast measurement with IOL calculation In single screen Printer |
| printer | Fast Thermal printer with Auto Cutter |
| Power | |
| Electrical Requirements | 230V AC / 50Hz |
| PACHYMETER Specifications 🙁 General ) | |
| Measurement Range | 200- 1000 microns |
| Accuracy | + or – 5 microns |
| Frequency | 20MHz |
| Communication | RS232 for PC connection |
IOP Adjust Software
- Easily Adjust IOP based on measured CCT.
- CCT measurements
- Guaranteed fast and easy Measurements
Scan Modes
- Continuous/Mapping
- User programmable Mapping (up to 10 maps)
Measurements
- Adjustable Corneal Velocity
- Automatic Sensing Algorithm
- Measurement Review
- 100 Scan Average with Standard Deviation
Probe
- 20 MHz Angled or Straight
Other Features
- Auto / Manual Gain
- Automatic calculation of stamdard deviation and average axial length ( serial of 10 measurement )
- Simultaneous display of 3 different IOL calculation
- RTC shown the Date Time throughout the life of the instrument.
- Various user Programmable printouts format
- Measurement possible for various velocity of each segment.
- *AME Formula – The new generation programmed software takes care of any Axial Length.